Назив: | Efficiency of whole-body counter for various body size calculated by MCNP5 software |
Аутори: | Krstic, Dragana Nikezic, Dragoslav |
Датум издавања: | 2012 |
Сажетак: | The efficiency of a whole-body counter for 137Cs and 40K was calculated using the MCNP5 code. The ORNL phantoms of a human body of different body sizes were applied in a sitting position in front of a detector. Theaim was to investigate the dependence of efficiency on the body size (age) and the detector position with respect to thebody and to estimate the accuracy of real measurements. The calculation work presented here is related to the NaI detector, which is available in the Serbian Whole-body Counter facility in Vinca Institute. © The Author 2012. Published by Oxford University Press. All rights reserved. |
URI: | https://scidar.kg.ac.rs/handle/123456789/10474 |
Тип: | article |
DOI: | 10.1093/rpd/ncs219 |
ISSN: | 0144-8420 |
SCOPUS: | 2-s2.0-84870461672 |
Налази се у колекцијама: | Faculty of Science, Kragujevac |