Молимо вас користите овај идентификатор за цитирање или овај линк до ове ставке: https://scidar.kg.ac.rs/handle/123456789/15701
Назив: ON THE PLACEMENT OF APRON DOSEMETERS AND DOSE ASSESSMENT IN INTERVENTIONAL CARDIOLOGY PROCEDURES: PRELIMINARY RESULTS
Аутори: Ferrari, Paolo
Becker F.
Campani L.
Jansen J.
Jovanović, Z.
Krstic, Dragana
Mariotti F.
Teles, Pedro
Venturi G.
Датум издавања: 2022
Сажетак: Personnel involved in interventional practices are likely to be exposed to higher radiation doses than other workers in the medical field. Personnel monitoring and radiation protection measures play a crucial role in keeping these doses below the limits. EURADOS (European Radiation Dosimetry Group) Working Group 12 performed a series of investigations showing how the complexity of the scattered field reaching the operators can influence the doses to the operators. The present work was aimed at determining the possible effects on the registered doses of the scattered field and the actual position of a dosemeter on apron. This study has been performed through Monte Carlo simulations and it was validated through measurements. It does not claim to identify the 'best' position for the dosemeter, but to assess the variability of its response, showing how a variability of the order of +/- 30% to 40 should be taken into account.
URI: https://scidar.kg.ac.rs/handle/123456789/15701
Тип: article
DOI: 10.1093/rpd/ncac188
ISSN: -
SCOPUS: 2-s2.0-85142400945
Налази се у колекцијама:Faculty of Science, Kragujevac

Број прегледа

378

Број преузимања

5

Датотеке у овој ставци:
Датотека Опис ВеличинаФормат 
PaperMissing.pdf
  Ограничен приступ
29.86 kBAdobe PDFСличица
Погледајте


Ставке на SCIDAR-у су заштићене ауторским правима, са свим правима задржаним, осим ако није другачије назначено.