Молимо вас користите овај идентификатор за цитирање или овај линк до ове ставке: https://scidar.kg.ac.rs/handle/123456789/18098
Назив: Contact stress generation on the UHMWPE tibial insert
Аутори: Petrovic Savic, Suzana
Adamovic, Dragan
Devedzic, Goran
Ristić, Branko
Matić, Aleksandar
Часопис: Tribology in Industry
Датум издавања: 2014
Сажетак: Total knee replacement (TKR) is considered, during last years, as a very successful surgical technique for removing knee joint deformities and eliminating pain caused by cartilage damage. In literature, as primary causes for knee joint endoprothesis damage are cited complex movements which cause occurrences of complex stress conditions, sagital radius conformity, sliding, types of materials etc. Aim of this study is analysis of contact stresses that occur on tibial implant for 15°, 45° and 60° knee flexion and 50 kg, 75 kg, 100 kg and 125 kg weight. Knee joint prosthesis model and finite elements method (FEM) analysis are done in software Catia V5. For this analysis we used ultra-high molecular weight polyethylene (UHMWPE) for tibial implant material and AISI 316, AISI 317, AISI 321, 17-4PH, CoCrMo, Ti6Al4V and SAE A-286 for femoral component materials. Results show that area of maximal contact stress is identified in medial and lateral part of tibial implant. Von Mises stress values vary regarding of flexion degree and weight, but values are approximate for types of chosen materials. Contact stress location corresponds to damage that occur on tibial implant during exploitation.
URI: https://scidar.kg.ac.rs/handle/123456789/18098
Тип: article
ISSN: 0354-8996
Налази се у колекцијама:Faculty of Engineering, Kragujevac

Број прегледа

47

Број преузимања

12

Датотеке у овој ставци:
Датотека Опис ВеличинаФормат 
Contact stress generation on the UHMWPE tibial insert.pdf689.39 kBAdobe PDFСличица
Погледајте


Ставке на SCIDAR-у су заштићене ауторским правима, са свим правима задржаним, осим ако није другачије назначено.