Молимо вас користите овај идентификатор за цитирање или овај линк до ове ставке: https://scidar.kg.ac.rs/handle/123456789/20614
Назив: Analysis of Experimentally Determined Transfer Function of Photoacoustic System Detection Part
Аутори: Todosijević, Slobodan
Ciric, Dejan
Radičević, Branko
Датум издавања: 2017
Сажетак: The microphone, amplifier and photoacoustic cell affect the transfer function of the gas-microphone detection of the photoacoustic measuring system. Experimental determination of the transfer function of the system detection part eliminates the influences of these components occurring due to the system design. Based on the measurements performed in an anechoic chamber where the swept sine signal was used as an excitation, the transfer function is determined. Two different acquisition cards were used for the signal detection. One of them was used to generate the excitation signal simultaneously with the microphone signal detection. The approximately flat transfer function up to 6 kHz is obtained with that card, which is not the case with the other acquisition card. In the latter case, the determined transfer function shows the oscillations in the regions where the signal to noise ratio is insufficient.
URI: https://scidar.kg.ac.rs/handle/123456789/20614
Тип: conferenceObject
Налази се у колекцијама:Faculty of Mechanical and Civil Engineering, Kraljevo

Број прегледа

343

Број преузимања

11

Датотеке у овој ставци:
Датотека Опис ВеличинаФормат 
2017_IcETRAN - Todosijević S. et al..pdf1.79 MBAdobe PDFСличица
Погледајте


Ставке на SCIDAR-у су заштићене ауторским правима, са свим правима задржаним, осим ако није другачије назначено.