Please use this identifier to cite or link to this item: https://scidar.kg.ac.rs/handle/123456789/12945
Title: Comparison of thyroglobulin concentrations measured by two immunoradiometric assay
Authors: Ignjatovic, Vesna
Matovic, Milovan
Vukomanović, Vladimir
Issue Date: 2020
Abstract: © 2020, University of Kragujevac, Faculty of Science. All rights reserved. Circulating thyroglobulin measurements is a highly specific test in the management of patients affected by differentiated thyroid cancer after total thyroidectomy, followed by radioiodine ablation. The aim of our study was to com-pare two thyroglobulinimmunoradiometric assays (INEP, Serbia and Cisbio Bioassays, France). Study included 42 patients of both genders with DTC. The subjects were on suppres¬sive doses of levothyroxine and followed up. Results showed concordance between the two assay methods for determining serum thyroglobulin for 39 (92.85%) pa-tients. Statistical analysis showed that there was a direct correlation between two IRMA tests, with a positive correlation coefficient r=0.613 (p 0.05). We concluded that there is good agreement between the two thyroglobulin assays compared in this study.
URI: https://scidar.kg.ac.rs/handle/123456789/12945
Type: article
DOI: 10.2478/SJECR-2018-0031
ISSN: 1820-8665
SCOPUS: 2-s2.0-85091198397
Appears in Collections:Faculty of Medical Sciences, Kragujevac

Page views(s)

540

Downloads(s)

21

Files in This Item:
File Description SizeFormat 
10.2478_sjecr-2018-0031.pdf176.76 kBAdobe PDFThumbnail
View/Open


This item is licensed under a Creative Commons License Creative Commons